AMS NB-201 Semi Auto Biochemistry Analyzer
FEATURES
- End point, kinetics, fixed time absorbance
- All-in-one optical system, high reliability and anti-interference
- 7 filters (340nm~630nm) + 1 free position
- Reaction curve and QC graph print-out
- Dynamic temperature control of Peltier elements 25°C, 30°C, 37°C
- Up to 112 test profiles can be programmed
- Memory for 3200 sample results
- Touch screen, large LCD display
SPECIFICATIONS
- Light source
- Quartz-halogen lamp 12V/20W
- Wavelength range
- Automatic by 8 positions filter-wheel
- 7 standard filters: 340nm, 405nm, 492nm, 510nm, 546nm, 578nm, 630nm
- 1 position free for optional filter
- Half bandwidth<8nm
- Stray light: <0,3% at 340nm
- Photometric range
- 0.0000 to 3.0000 ABS
- Resolution: 0.0001 ABS
- Stability: 0.002 A/hour
- Flowcell
- Stainless steel with quartz window measuring volume 32ml
- Optical path 10mm
- Aspiration volume programmable; 200~3000ml
- Carryover<1.0%
- Analysis method
- End point, with or without reagent blank
- End point, with sample bland, with or without reagent blank
- Kinetic, with or without linearity check
- Two-point kinetic/fixed time, with our without reagent blank
- Absorbance
- Turbidimetry
- Linear and non-linear calibration
- Parameter settings
- Method
- Wavelength
- Temperature
- Reagent blank y/n
- Sample blank y/n
- Lag time
- Measurement time
- Reaction time
- Absorbance limit
- Aspiration volume
- Standards
- Linearity check
- Unit for results
- Memory
- 112 test profiles
- 3200 sample results
- Temperature control
- By means of Peltier elements
- 25°C, 30°C, 37°C ± 0.1°C
- Input
- Touch screen
- Display
- Large LCD display
- Print
- Built-in thermal recorder
- Interface
- RS-232 serial port
- Operating environment
- Temperature: 10°C~30°C
- Humidity: ≤70%
- POWER REQUIREMENT
- AC 220V/110V ± 10%, 50Hz ± 2%
- Dimension
- 420mm (H) x 380mm (W) x 170mm (D)
- Weight
- 8.5kg
Reviews